By AZoNano
Topics Covered
Background
Correlating AFM
Surface Potential Measurements In Air With Open-Circuit Electrochemical
Potential Measurements In Electrolytes
AFM Provides
High-Resolution Surface Potential Measurements
Comparing AFM Surface
Potential Imaging With Electron Based Probe Techniques
Summary
Other SPM
Techniques for Corrosion Science Research
Background
Atomic Force Microscopy
(AFM) offers 3 primary modes for imaging:
- Contact Mode
- TappingMode
- Torsion
- Resonance Mode (TRmode)
Each primary mode enables
numerous other modes, which we collectively refer to as secondary modes
or derivative modes. Surface potential imaging, or Scanning Kelvin
Probe Force Microscopy is a derivative of TappingMode AFM that is
described in detail elsewhere. What makes surface potential imaging
different from most other derivative AFM modes is that, it furnishes
reliable, repeatable numerical values of a quantity other than
topographic dimensions, and maps that quantity across the sample
surface area, simultaneously with topography. That quantity is the
electrostatic potential of a small area—immediately underneath the AFM
tip—on the sample surface, and it is measured relative to the potential
of the AFM tip.
This application note
first compares AFM-based surface potential measurements with
electrochemical potential measurements made in bulk electrolytes, and
shows the correlation. It then shows how the information in AFM surface
potential images and electron micrographs of the same area of a sample
surface may complement each other.
The application note also
shows how AFM surface potential measurements and electron scattering
data correlate. In short, the images and data presented here help
establish the value and usefulness of AFM in corrosion science
research, by illustrating that the qualitative and quantitative results
stemming from the unique capabilities of AFM surface potential
measurements and imaging correlate with results from other analytical
techniques.
Correlating AFM Surface Potential Measurements In Air
With Open-Circuit Electrochemical Potential Measurements In Electrolytes
In order to establish
that the quantitative output of AFM-based surface potential measurement
is meaningful for corrosion studies, in the sense that it correlates
with measurements made with other techniques, we compared AFM-based
surface potential measurements in air with electrochemical potential
measurements in bulk electrolytes, as described next.
Samples of different
metals were immersed for 30 minutes in either de-ionized water (DI-H2O) or a 0.5M
aqueous NaCl solution. The stabilized open circuit potential was then
measured for these samples versus a reference saturated Calomel
electrode (SCE) using a potentiostat. The samples were then removed
from the electrolytes, rinsed with DI-H2O, and air-dried prior
to AFM surface potential imaging and measurements with a Dimension
SPM. The AFM tip
was coated with a metal layer. The potential of the tip was calibrated
by measuring the surface of a pure Ni sample after exposure to water;
this surface was found to provide reproducible measurements.
The AFM-based surface
potential measurements are plotted in Figure 1, along the left-right
axes, versus the open circuit bulk electrolyte measurements for DI-H2O, and for the
NaCl solution. The AFM measurements correlate linearly with the open
circuit potentials, and can be associated with the Volta potential of
the samples. The open circuit potentials measured in 0.5M NaCl are
shifted in the active direction (i.e., they are lower) by around 200mV
compared with measurements made in DI-H2O.

Figure 1.
Correlation of AFM Surface Potential Measurements in air (“Volta potential”)
with open circuit potential measurements in electrolytes for different
metals A) in de-ionized water, B) in 0.5M NaCl. Reproduced by
permission of The Electrochemical Society.
These plots establish
that surface potential measurements made with the AFM are reliable for
establishing the relative nobility or activity of species, and can be
compared with open-circuit potential measurements made of the same
species in bulk electrolytes. The main difference is that AFM surface
potential measurements may have very high in-plane (X,Y) resolution,
allowing for deep sub-micrometer mapping of regions with different
potentials, as we describe next.
AFM Provides High-Resolution Surface Potential
Measurements
A unique feature of AFM
surface potential imaging is that it maps the potential locally, with a
resolution that can extend down to the nanometer-scale in the plane of
the sample surface. Figure 2 shows the surface of an aluminum-based
AA2024-T3 alloy, commonly used in airplanes and extremely susceptible
to corrosion. The left and right images are TappingMode AFM topography
and surface potential maps, respectively, of the same area, 60ìm on a
side.

Figure 2.
AFM Images of an AA2024-T3 alloy sample. Inter-metallic particles are
visible as brighter areas (higher potentials) in the surface potential
image (right). Topography (left) does not distinguish between the
matrix and the inter-metallic particles. 60µm scans. Reproduced by
permission of The Electrochemical Society.
Lighter shades of color
in surface potential images correspond to higher potential values. On
this sample, inter-metallic particles that are not evident in the
surface topography are seen in sharp contrast to the embedding matrix
in the surface potential image, which is captured simultaneously with
the topography map. The surface potential image pinpoints the location
and boundaries of inter-metallic particles. These particles measure
from sub-micrometer to as large as 20ìm across, and here they exhibit a
higher potential than the alloy matrix. These potential differences are
responsible for enhanced corrosion due to galvanic coupling between the
different areas. The cathodic reaction rates are enhanced at these
particles, while dissolution is stimulated at lower potential sites in
the matrix or at active particles.
The surface potential
image can also shed some light on the significance of the features in
the TappingMode AFM topography image. In Figure 2, several pits are
visible in the topography image (left): one associated with the
inter-metallic particle 5 in the surface potential image, and one with
particle A. This sample’s surface was prepared by non-aqueous polishing
to minimize corrosion during preparation, and was examined in the
as-polished condition. The pits evident in the topography image formed
during polishing (despite the non-aqueous polishing medium), because
the sample is extremely susceptiable to corrosion during the polishing
process. The measured potential distribution (right) provides important
information regarding the most likely locations for the anodic and
cathodic reactions on the surface during subsequent exposure to a
corrosive environment.
Comparing AFM Surface Potential Imaging With Electron
Based Probe Techniques
Figure 3 is a scanning
electron micrograph (SEM) of approximately the same area shown in the
AFM images in Figure 2. The contrast in the SEM results from the
difference in the electron-scattering properties between the particles
and the matrix. EDS analysis performed on different particles confirmed
that the regions of higher potential in Figure 2 are associated with
different types of inter-metallic particles: particles 1-5 are
Al-Cu-(Fe,Mn), and particles A and B are Al-Cu-Mg. In the surface
potential image in Figure 2, particles 4 and 5 measure at lower
potentials than particles 1-3; so, there are apparently different types
of Al-Cu-(Fe,Mn) particles present in the matrix.

Figure 3.
SEM image of approximately the same area as in Figure 2. EDS analysis
indicated that particles 1-5 are Al-Cu-(Fe,Mn) inter-metallics, and
that A, B are Al-Cu-Mg inter-metallics. Reproduced by permission of The
Electrochemical Society.
These are further
evidence that surface potential images can complement information
obtained from established analytical techniques, and vice-versa. AFM
surface potential imaging reveals effect of processing step Experiments
have shown that the Mg containing particles will dissolve upon exposure
of the polished surface to a chloride-containing solution, but that the
attack takes some time to initiate. During this initiation time, the
potential of these particles decreases to the value of the potential of
the matrix.
Surface Potential imaging
is extremely sensitive to surface charge. The as-polished AA2024-T3
samples are covered by a native oxide layer formed during polishing and
subsequent exposure to air. We imaged an area of a sample surface, and
identified in that area several round Mg-containing particles before
and after removing some of the oxide with Argon ion sputtering (Figure
4). Auger analysis performed simultaneously with the sputtering showed
that the sputtering removed only part (1-2nm) of the surface oxide
thickness. The sample was then exposed to air again and the surface
potential was remapped with AFM. The surface potential images in Figure
4 show that after removing a few mono-layers of the oxide from the
surface, the measured potential of the Mg-containing particles shifted
from being more noble than the matrix (higher potential, thus brighter
colors in the image), to being more active (darker colors). The
location and the boundaries of these particles are visible with
sub-micrometer resolution.

Figure 4.
Surface Potential images of inter-metallic particles in AA2024-T3. The
arrows point to the location of Mg-containing particles. (Left) before,
and (right) after partially removing (about 1-2nm) the native oxide
film by argon ion sputtering. 30ìm scans. Reproduced by permission of
The Electrochemical Society.
Surface potential
measurements using data histograms (Figure 5) quantify the shift in
potential to be from about 60mV higher than the matrix to about 60mV
lower. When this sample was exposed to a chloride solution, these
particles dissolved immediately with no initiation time required. The
redistribution of charge during the partial removal of the surface
oxide film and re-growth in air resulted in activation of the
particles, and this had an immediate consequence for the corrosion
behavior.

Figure 5.
Histograms of surface potential images in Figure 4. The round
Mg-containing particles (correlating approximately to the position of
red cursor in histograms) shifted from being about 60mV higher in
potential than the matrix (green cursor in histograms), to being about
60mV lower after partial native-oxide removal (bottom histogram). (The
peak on the right side in both histograms corresponds to the large,
irregularly shaped particle, which is the dominant feature at the
center of both images.
Summary
AFM-based surface
potential imaging and measurements are able to reveal details on the
surface of a sample in unique ways that are useful to corrosion science
research. These images and measurements, often with nanometer-scale
in-plane resolution, are complementary to, and correlate with, data
from other analytical techniques, including bulk techniques.
Other SPM Techniques for Corrosion Science Research
Soon after the
development of AFM and Scanning Tunneling Microscopy (STM),
electrochemical environment control of the sample was also introduced,
in the form of open or closed liquid electrolyte cells, with
electrochemical cell potential control and voltammetry display and
analysis software integrated with the AFM (and STM) software. Today,
electrochemical AFM and STM, play an important role in corrosion
science research.

This information has been sourced, reviewed and adapted from
materials provided by Bruker AXS.
For more information on this source please visit Bruker AXS.