Bimodal Dual AC Imaging from Asylum Research

Topics Covered

Background
Bimodal Dual AC Imaging-How it Works
Feedback
Bimodal Image Examples
Graphite
DNA in Fluid
Conclusion

Background

Dual AC imaging encompasses a wide variety of techniques. We have broken down these various imaging techniques to specifically describe the mode for different applications. Table 1 explains the various Dual AC techniques along with the MFP-3D software settings needed for imaging. For this application note, only bimodal Dual AC imaging will be discussed.

 

 

MFP-3D Software Settings

Technique

Description

Frequency 1-f1

Frequency 2-f2

Drive 1-2

Mode: Drive

Bimodal

Drive the 1st and 2nd resonance.

At or near fundamental resonance

At or near 2nd resonance

On-On

Dual AC: Shake

MFM

Bimodal with special operating parameters and a magnetic cantilever.

At or near fundamental resonance

At or near 2nd resonance

On-On

Dual AC: Shake

Passive bimodal

Similar to bimodal, but the 2nd resonance is only monitored, not driven.

At or near fundamental resonance

At or near 2nd resonance

On-Off

Dual AC: Shake

Active harmonic

Similar to bimodal, but a higher harmonic driven.

At or near fundamental resonance

f2=Nxf1 where N is an integer

On-On

Dual AC: Shake

Passive harmonic

Higher harmonic is simply monitored, not driven.

At or near fundamental resonance

f2=Nxf1 where N is an integer

On-Off

Dual AC: Shake

DFRT

By driving at one frequency below resonance (A1) and another above (A2), the A2-A1 gives an error signal we can use to track the resonance frequency changes.

Slightly below fundamental resonance (A~Amax/2)

Slightly above fundamental resonance (A~Amax/2)

On-On

Dual AC to Shake

DFRT-PFM

Piezo Response Force Microscopy technique using the above.

Slightly below fundamental resonance (A~Amax/2)

Slightly above fundamental resonance (A~Amax/2)

On-On

PFM: Dual AC to Chip

Bimodal Dual AC Imaging-How it Works

Bimodal Dual AC imaging takes advantage of the flexibility and power of the digital signal processing within the MFP-3D AFM system controller. The idea behind this is simple. Cantilevers are extended mechanical objects and have many different flexural resonant frequencies. In the past, AC atomic force microscopes have typically excited one of those modes, usually the lowest frequency, or "fundamental" mode and then used the amplitude or frequency of that mode as the input for a feedback system that controls the tip-sample separation. When the amplitude of the fundamental motion is used, the term "Amplitude Modulated AFM" (AM-AFM) has been employed.

Figure 1. Bimodal Dual AC 2nd mode amplitude overlaid on rendered AFM topography (top) and fundamental phase image overlaid on topography (bottom) of multi-component surf wax. Notice the high contrast visible in the bimodal Dual AC image and the relative lack of contrast in the fundamental phase. 4µm scan.

A plucked guitar string has higher flexural resonances that are harmonic, meaning that if the fundamental mode is at a frequency f0, the next resonance is at 2f0, the next is at 3f0 and so on. For all but very a few very specialized types, cantilever resonant frequencies are non-harmonic. Butt and Jaschke as well as Sarid give nice reviews of cantilever mechanics. Figure 2 gives the theoretical resonant frequencies for diving board shaped cantilevers for the first few resonant frequencies. The measured resonant frequencies for two representative cantilevers, the Olympus AC-240 (for air imaging) and Olympus Bio-Lever (for fluid imaging) are also listed.

Figure 2. Theoretical resonant frequencies of diving board shaped cantilevers in terms of the fundamental, w1, as well as resonant frequencies for the Olympus AC240 and Bio-Levers. In these cases, the theoretical predictions typically agree with the measured results to well within 10%.

One of the simplest methods for looking at the effect of higher modes is to drive them directly, using the same machinery set-up as that used in AC imaging. As with AC imaging using the first resonant frequency, the amplitude of the cantilever measured at the drive frequency is used as the error signal in a feedback loop. Stark et. al. observed enhanced phase contrast on a sample using the third resonant frequency of a triangular cantilever. Crittenden et al. have also explored using higher harmonics for imaging because the response of the higher harmonics is sharper than the response of the fundamental resonance.

If the periodic repulsive interactions between the tip and sample are nonlinear, they will couple motion into higher harmonics. This can provide information about the mechanical properties of the sample. During a single AM-AFM vibrational cycle, the tip typically samples a range of forces, from the long range attractive to the short range repulsive. If the tip interacts with the short ranged repulsive forces, information about the mechanical properties of the sample can be obtained.

For single oscillation mode imaging, if the phase shift is positive, it is customary to refer to the imaging mode as "net attractive" or simply "attractive". If the phase shift is negative, the mode is referred to as "repulsive". Recently, Rodriguez and Garcia published a theoretical simulation of a non-contact, attractive mode technique where the cantilever was driven at its two lowest resonant frequencies. In their simulations, they observed that the phase of the second mode had a strong dependence on the Hamaker constant of the material being imaged, implying that this technique could be used to extract chemical information about the surfaces being imaged. Other work by Garcia's group has also demonstrated that the compositional sensitivity of an AFM is enhanced by the simultaneous excitation of its first two normal resonant frequencies 1-2.

Figure 3 shows the basic idea of bimodal Dual AC imaging mode using two typically non-harmonic resonant frequencies of a cantilever. The cantilever is driven with a linear combination of sinusoidal voltages, at or near the resonant frequencies, f1 and f2. This signal is used to drive the base of a cantilever with a "shake" piezo. The experiments reported here were repeated with a magnetically activated cantilever with similar results. It is expected that other actuation methods where two drive waveforms can be summed will prove as effective. The resulting motion of the cantilever is measured with a position sensor. This signal in turn is used as the input for two separate lock-in amplifiers, where the function generator f1 is used as a reference for one lockin and f2 is used as a reference for the other. The output of the lockin amplifiers, including the Cartesian in-phase and quadrature pairs (x1, y1, x2, y2) and polar amplitude and phase (A1, Ø1, A2, Ø2) representations of the cantilever motion at the two or more frequencies can then be passed on to the controller where they can be displayed, saved, combined with other signals, and used in feedback loops.

Figure 3. In bimodal Dual AC, the cantilever is both driven and measured at two (or more) frequencies. The sinusoidal "shake" voltage is a sum of voltages at frequencies f1 and f2. The cantilever deflection then contains information at both of those frequencies, as shown in the red curve. The amplitude and phase at the two frequencies are then separated again by the two lockins and passed on to the controller. The controller can use one or both of the resonant frequencies to operate a feedback loop.

Feedback

As with conventional AC imaging, the amplitude of the cantilever is used as the feedback error signal. There is a difference here, however, since there are two amplitudes – one at each drive frequency. The initial results we present use the amplitude of the fundamental frequency A1 as the feedback error signal and fundamental phase Ø1, the second resonant frequency amplitude A2 and phase Ø2 as "carry-along" signals. Reversing this and using the higher resonant frequency amplitude as a feedback and carrying the fundamental amplitude and phase along can also yield interesting results. The sum of all of the amplitudes as the error signal also allowed stable imaging.

An interesting feature of this measurement is that the signal processing can be performed on the same cantilever deflection data stream for each flexural mode. With a digital lockin implementation, for example, this implies that the same position sensitive detector and analog to digital converter (as long as it has sufficient bandwidth for the higher mode) can be used to extract information regarding the distinct resonant frequencies.

Bimodal Image Examples

Graphite

Figure 4 shows a 30µm image made on a highly oriented pyrolitic graphite (HOPG) surface. The cantilever was a silicon AC-240 cantilever from Olympus. It was driven at the fundamental (f1~69.5kHz, A1~8nm) and second resonant frequency (f2~405kHz, A2~8nm). No significant differences were observed for similar cantilevers imaging the graphite surface. The Z-feedback loop was operated using the fundamental amplitude A1 as the error signal. The topography (A) shows the expected terraces separated by single or multiple atomic steps. The first mode amplitude (B) channel resembles a high-pass filtered image of the topography. The fundamental phase image (C) shows an average phase lag of ~34° and very little variation (≤1° standard deviation), implying that the cantilever was consistently in repulsive mode. Again, there is very little contrast in the fundamental phase image. The second mode amplitude image (D) however has significant contrast, with broad patches showing regions where A2, the second mode amplitude, was reduced by tip-sample interactions. A three dimensional rendering of the surface topography (A) with the second resonant frequency amplitude (D) "painted" onto the rendered surface (E) allows the high contrast second mode data to be correlated with the topography. Although (E) makes it clear there is a high degree of correlation, there are also boundaries in the second mode amplitude that seem to have no connection to topographical features.

Figure 4. Graphite images A-D showing different imaging modes (top) and bimodal Dual AC 2nd mode amplitude overlaid on AFM rendered topography (bottom), 30µm scan.

DNA in Fluid

Bimodal Dual AC imaging also works well for AM-AFM imaging in fluids. A high density Λ-digest deoxyribonucleic acid (DNA) sample was prepared in a dense mat on freshly cleaved mica. Figure 5 shows the response of a 60µm long Olympus Bio-Lever in fluid being driven at its fundamental resonance (f1~8.5kHz, A1~8nm) and at its second mode (f2~55kHz, A2~5nm) in the DNA buffer solution. The topography (A) shows a dense mat of material on the surface with no clear strands of DNA visible. Similarly, the fundamental amplitude (B), the channel used for the feedback error signal, shows no particular structure. The fundamental phase channel (C) shows subtle contrast between the background and a structure that shows hints of being strands of DNA molecules. The second mode amplitude (D) shows clear, high contrast images of the same DNA strands. The strands appear dark, corresponding to an increased dissipation. This is consistent with the DNA strands being slightly less bound to the sample and thus able to absorb some of the second resonant frequency energy. Again, rendering the topography in three dimensions and painting the second mode amplitude on top (E) allowed the topography and second mode amplitude to be spatially correlated.

Figure 5. (A) topography, (B) fundamental amplitude, (C) fundamental phase, (D) bimodal Dual AC second mode amplitude of DNA, 750nm scan. (E) Second mode amplitude data overlaid on rendered AFM topography.

Conclusion

By measuring the cantilever response at two different frequencies, it is possible to look at the difference in, for example, the phase signals at the fundamental drive frequency and at a higher mode drive frequency. In the future this may help with extracting frequency-dependent mechanical properties of the sample. Significant contrast differences can be observed by operating a repulsive mode AM-AFM cantilever at more than one of its flexural resonances. As research continues, we're constantly learning more about Bimodal Dual AC. Contact Asylum Research to learn about the latest discoveries and applications on this technique.

Source: Bimodal Dual AC™ Imaging
A complete set of references can be found by referring to the source document
For more information on this source please visit Asylum Research

Date Added: May 7, 2008 | Updated: Jun 11, 2013