.jpg)
Topics Covered
Introduction
Integration of Atomic Force Microscope and Inverted Light
Microscope
CoverslipHolder and BioCell Designed to Optimise Image Quality
Piezos in JPK Instruments
Testing the Calibration of Optical Images
Conclusion
Introduction
Atomic force microscopy (AFM) and optical microscopy, in
particular fluorescence microscopy, make a powerful
combination in the study of biological samples. AFM is not
subject to Abbe's resolution limit, and can generate images
with a much higher resolution than light microscopy.
However, as contrast is generated in response to the
structural properties of the sample, it can be challenging to
detect specific structures in a heterogeneous sample, such
as a cell. By combining the two techniques, higher
resolution structural information can be generated using
AFM. Subsequent correlation with fluorescently labelled
markers can provide information about the composition,
and consequently the function, of the identified structures.
Integration of Atomic Force Microscope and Inverted Light Microscope
The design of the NanoWizard®
and NanoWizard®II
AFM from JPK
is such that the atomic force microscope is integrated into an inverted light
microscope, without affecting its functionality. This allows AFM imaging to
be combined with contrast techniques including phase contrast, DIC, laser scanning
confocal and TIRF microscopy. An important factor in the efficacy of this integration
is that the JPK
instruments are tip-scanners. That is, during AFM imaging the sample is
held still, while the tip moves in raster fashion over the surface to build
the image. In the case of sample-scanning instruments, the sample to be imaged
is constantly moving in the optical microscopy picture while AFM scanning is
in progress, meaning that true simultaneous imaging is not really possible (figure
1).
.jpg)
Figure 1. Tip scanner vs sample scanner. The
acquisition of epifluorescent images during AFM image acquisition with a tip
scanner (A) and a sample scanner (B). As the sample is moving in (B) the fluorescent
structures can no longer be imaged without smearing. The same cell is shown
in both images.
CoverslipHolder and BioCell Designed to Optimise Image Quality
While these hardware design features are the start of
providing true integration there are also other factors that
need to be addressed. The first is the sample holder used.
As AFM and optical microscopy are fundamentally different
techniques (one being based on physical interaction and
the other on the diffraction of light) there are different
requirements for effective sample supports. For the
acquisition of high quality optical images, particularly with
high magnification lenses (63x and 100x) it is best to use
very thin cover-glass with a thickness of around 170 µm.
On the other hand, as AFM imaging is very sensitive to physical instability,
the support for such imaging must be very stable. With these two fundamental
requirements in mind, JPK
designed the CoverslipHolder
and the BioCell™
(Figure 2). Both of these sample holders are designed to hold coverslips for
uncompromised, combined AFM and light microscopy imaging. As such, with innovative
sample holders and the fundamental design of the JPK
instruments, simultaneous, high quality AFM and light microscopy images
can be acquired. However, to be truly integrated we have now introduced the
DirectOverlay™ feature (patent pending), a software solution that involves
calibration of the optical image and integration into the SPM software.
.jpg)
Figure 2. The Biocell™ is designed to
optimise image quality during experiments combining AFM and light microscopy
while at the same time allowing temperature control. Peltier elements allow
rapid adjustment of temperature.
Piezos in JPK Instruments
As optical microscopy is based on the use of lenses, any aberrations in such
lenses will lead to distortions in the final image. However, as the piezos in
the JPK
instruments are linearized the AFM image is precise to 4Å in the x
and y directions. As such, in most cases the AFM image and the light microscopy
image do not accurately overlay, with shear or stretch in the optical image
a common problem.
As the AFM image is generated using very precise linearized piezos it can be
treated as "real-space". Additionally, as the cantilever used for
AFM imaging can be moved to fixed points, as well as raster-scanned over the
surface, it can be used to calibrate the optical image. In short, the cantilever
is moved to a set of 25 points in realspace, using the piezos. At each point
an optical image is acquired and subsequently the tip location within the optical
image is automatically determined. A transform function is then calculated using
both sets of 25 points, and this transform is applied to the optical image as
it is imported into the SPM software. In such a way the optical image is calibrated
and imported into the SPM environment, in an automated process.
Testing the Calibration of Optical Images
In order to test the calibration of the optical image, 50 nm
fluorescent beads were imaged using both AFM and epifluorescence
and both the untransformed and the
transformed optical images compared with the AFM. It can
be seen that there is a shear in the right hand side of the
untransformed optical image (figure 3). Once the
calibration procedure has been applied, however, the
overlay between the fluorescent and topographic signals is
precise (figure 4).
.jpg)
Figure 3.Overlay of topography (red) and untransformed
fluorescence (green) images of 50nm beads. The lower panel is a set of digital
zooms of the marked regions. While in the first two of the zoomed regions the
overlay is good, in panel 3 there is a shear of the optical image.
.jpg)
Figure 4.Overlay of topography (red) and untransformed
fluorescence (green) images of 50nm beads. The same area that was imaged in
Figure 3 is shown here. In all zoomed regions the overlay is now precise.
The benefits of such a software feature are extensive. For instance, in the
comparison of fluorescence and AFM images there may not be easily identified
fixed points within the two images to conduct such a transformation offline,
nor even to overlay the edges. As seen in figure 5, the overlay of the AFM image
of a REF52 fibroblast (stably transfected with YFP-paxillin) and the corresponding
fluorescence image of the focal adhesions, there are no points in both images
that can easily be identified for use in accurately mapping one image to the
other.
.jpg)
Figure 5.Overlay of fluorescence and topography
of REF52 fibroblast cells. In panel A is the transformed fluorescence image
and B topography. In C an overlay of the two is displayed.
In this case the focal adhesions are at the basal side the cell and the AFM
image generates a topograph of the apical side of the cell. Thus the use of
the cantilever as a tool for calibrating the optical image is essential. In
figure 6 the difference between the transformed and non-transformed optical
image is displayed. In this case the predominant artefact is a stretch along
one axis.
.jpg)
Figure 6.Overlay of an image that has been
calibrated (red) and the same image in its raw, uncalibrated form (green) is
presented. It can clearly be seen that the uncalibrated image is stretched in
one direction.
Additionally, such a feature can save the user a considerable amount of time.
While AFM has high spatial resolution, the temporal resolution of the technique
is far lower than that of light microscopy, due to the longer acquisition times.
With a calibrated optical image in the back ground of the SPM software (figure
7), the need to acquire an overview AFM image of the cell before focussing on
a region of interest is removed. This can be particularly important for the
scanning of regions of a living cell, where time may be important. Obviously,
forcespectroscopy points can also be selected on the optical image, again removing
the need to acquire an AFM image before starting force-spectroscopy experiments.
When a functionalized tip is being used this could prove critical, as if a scan
is taken before the force-spectroscopy data is obtained the tip may be passivated,
leading to false negative results.
.jpg)
Figure 7.Import of optical images into SPM
software. A calibrated optical image can be imported into the SPM software and
displayed in the background. This allows the selection of scan regions and force-spectroscopy
points based on the optical image, removing the need for an overview AFM image.
Conclusion
The alterations in atomic force microscope design that led to it's installation
on an inverted light microscope opened the possibility for the simultaneous
acquisition of light and AFM images, crucial for the effective investigation
of cells and biological systems in vitro. However for true optical integration
with AFM, more is required than just the colocalisation of the two microscopes.
A tip scanner is essential such that the sample is still during AFM imaging,
so that optical images can be acquired. Sample holders must be optimised for
both forms of microscopy, i.e. thin coverslips for light microscopy and stability
for AFM imaging. Finally, a calibrated optical image must be available in the
AFM imaging software, to enable accurate overlays, with artefacts from aberrations
in the optical image removed. The NanoWizard®II
AFM provides all of these features, allowing for the first time true integration
of optics and AFM.
Source: JPK
Instruments
For more information on this source please visit JPK
Instruments