.jpg)
Topics Covered
Background
Introduction
Instrumentation
Experimental Results and
Discussion
Example 1: UHMWPE
Orthopaedic Bearings and
Materials
Summary
Example 2:
Drug Coated Cardiovascular
Stents
Summary
Example 3:
Vestamid and Pebax
Catheters
Summary
Example 4:
Contact
Lenses
Summary
Conclusions
Background
Materials property characterization is central to nearly all facets of
medical device design, production, and pre or post-use testing. Material
properties are routinely evaluated to meet needs in R&D & quality
assurance, and to obtain regulatory approval for sale. Devices that have been
used clinically or experimentally are also routinely analyzed to better
understand mechanisms of degradation, wear and tear, and/or failure.
Polymers are widely used either solely or in combination with other materials
in applications that include catheters, electrode lead insulations (e.g. cardiac
pacing or nerve stimulating), orthopaedic bearings, sutures, drug release
coatings (e.g. cardiovascular stents) and many more. Polymers are heterogeneous
materials that can have substantial variations in physical properties over
micron to millimeter length scales due to changes in molecular weight,
crystallinity, mixing uniformity of drugs in pharmacologically active coatings,
and the degree and morphology of phase separation in blends, co-polymers, and
composites.
Morphological characterization of polymers is typically done with different
types of microscopy, while functional and intrinsic properties such as molecular
weight (MW) and mechanical strength are typically evaluated on bulk specimens
without providing spatial information. Bulk thermal analysis techniques
including differential scanning calorimetry (DSC), thermal mechanical analysis
(TMA) and others are also widely used to elicit information on polymer
properties such as phase and component mixing.
However, these thermal analysis methods are limited to the assessment of the
average or aggregate properties of samples that are hoped to be "representative"
of manufactured devices, since such samples are necessarily removed from or
created separately from manufactured devices. Since "representative" analytical
specimens are required, in situ variations in material properties on
manufactured devices may be missed. In addition to limitations on the analysis
of manufactured devices and regions of devices, existing thermal analyses are
also nearly impossible to perform on thin polymer coatings.
Introduction
The Vesta
(Anasys Instruments) is a new analytical instrument that enables
manufactured medical devices to be analyzed for functional thermal transition
properties via Transition Temperature Microscopy (TTM). With TTM, the assessment
of thermal properties may be made in any identified specific region of a device
or specimen, thus uniquely integrating spatial information with thermal
properties. This thermal property measurement provides information on MW,
crystallinity, component mixing, and phase segregation that can be used to guide
R&D, enable production methods development, and provide production quality
control. Additionally, these thermo mechanical measurements also provide insight
into mechanisms of localized degradation, wear and/or failure of devices after
testing or following clinical use.
This application note illustrates the assessment of thermal transition
properties of several different types of polymeric medical devices and discusses
the information that can be gleaned from such analyses.
Instrumentation
The Vesta
Instrument incorporates a micro-machined inverted pyramidal thermal probe
that is fabricated on an AFM-like (atomic force microscope) cantilever. An
optical microscope enables users to position the thermal probe onto sample
regions of interest. When directed, the Vesta system will then bring the <30
nm radius thermal probe tip into mechanical contact with the specimen.
The probe can also be programmed for automatic positioning in order to
perform a series of measurements or to generate a Transition Temperature Map
(TTM). Once in contact with the region of interest the probe tip is resistively
heated while the cantilever provides sensitive force control and measurement
between the tip and sample. This provides localized thermo mechanical analyses
that are displayed in the form of a plot of the physical deflection of the
thermal probe on the specimen versus the probe temperature.
With most specimens (e.g. Figure 1), as the probe is heated the specimen
first expands and thus moves the probe tip up. At a transition temperature, the
specimen will begin to soften and the tip thus begins to move down into the
specimen. This temperature versus deflection curve is a local thermal analysis
(LTA) plot. The probe can be heated up to 450°C at ramp rates up to
600,000°C/min, thereby enabling high throughput measurements at temperatures
appropriate for essentially any polymeric or organic material. Material
softening, melting and even remelting can be monitored to provide information on
MW, phase and blend mixing and other physicochemical properties. The spatial
resolution of the measurement is dependent on the thermal probe radius of 30 nm,
and on the local thermal properties of the material. With most polymeric
systems, this enables local thermal analysis at a sub 100-nm scale.
Experimental Results and Discussion
Example 1: UHMWPE Orthopaedic Bearings and Materials
Ultra high molecular weight polyethylene (UHMWPE or PE) is an extremely
successful bearing material for orthopaedic hip and knee bearings, with millions
of devices in use and about a hundred thousand new PE-bearing hip and knee
arthroplasty devices implanted per year. Nonetheless, it has been long
established that wear of the PE impairs device function and that wear
particulates released from these bearings generate chronic inflammatory
responses.
Consequently, there has been extensive research to improve device durability
that has revealed mechanisms that include mechanical wear and oxidation that are
correlated with changes in PE molecular weight, crystallinity, and cross-link
density. Although it is not entirely clear what material properties are
important for long term biocompatibility in total hip and knee arthroplasty
prosthetics, physical cross-link density and crystalline morphology are two
properties that are well known to affect PE bearing function.
The evaluation of PE crystalline structure is difficult and laborious,
requiring tedious sample preparation for transmission electron microscopy in a
process that requires several days with specimens that are removed from bulk
samples or devices. Until now, there have been no rapid analysis methods for
crystallinity (and cross-link density), and no methods to enable the direct
evaluation of these critical properties in manufactured devices. The Anasys
Vesta, as an instrument that evaluates thermal transition properties that
are directly dependent on MW, crystallinity and cross-link density, uniquely
enables rapid measures in both experimental materials and in production
devices.
In the present study, several UHMWPE’s were analyzed including conventional
UHMWPE’s as in wide use, radiation cross-linked UHMWPE’s since these appear
tougher and less prone to wear and UHMWPE’s infused with alpha-tocopherol
(vitamin E) provided as an antioxidant. In addition, a retrieved explanted knee
bearing was examined demonstrating molecular degradation in identified
regions.
Figure 1 shows several representative local thermal analysis (LTA) plots for
virgin UHMWPE (GUR 1050 resin) with a viscosity average MW of approximately 5 x
106. This figure also shows LTA plots for the same PE after
cross-linking with 100 kGy radiation. The un-crosslinked PE shows a smooth curve
that is characteristic of MW and compositional uniformity, with a
Tm onset at 139°C. In contrast, the LTA plots of the
radiation cross-linked PE exhibited substantial irregularities indicative of
molecular heterogeneity. The Tm of this material occurred at 262°C, along with a
shoulder at 139°C that is consistent with DSC analysis of this material. [4]
Additionally, the probe penetrated less than half as deep into the radiation
cross-linked PE than the un-crosslinked PE. These thermal mechanical properties
are consistent with known and expected properties of radiation cross-linked PE.
.jpg)
Figure 1. Local thermal analysis (LTA) plots show
mechanical deformation of un-crosslinked UHMWPE (smooth blue curves peaking at
~139°C) and 100 kGy radiation cross-linked UHMWPE (red, less smooth, curves
peaking at ~262°C). Inset photo shows thermal probe cantilever overlying the
un-crosslinked specimen, with the arrows indicating the location of the thermal
probe tip that is under the cantilever.
Figure 2 shows three LTA plots for an alphatocopherol (vitamin E)
impregnated sample of the radiation crosslinked PE. The sample has a gradient of
alpha-tocopherol content that is clearly visible (not shown). The LTA curves
taken at different points across the gradient show a substantial effect on the
thermal transition curves, with higher levels of alphatocopherol correlating to
a softer material at 120°C, which is below the Tm. Thus,
alphatocopherol induces a concentration-dependant plasticization
effect.
.jpg)
Figure 2. Local thermal analysis plots of alphatocopherol
impregnated radiation crosslinked UHMPE. Softening of the material is observed
to correlate with alpha-tocopherol content.
An explanted UHMWPE knee bearing was examined that showed substantial
material breakdown, especially on the left side of the photo (3a). The detailed
history of this clinical implant is not known, but this implant was produced
about two decades ago. To evaluate the capability of the Vesta to assess
material properties in less obviously damaged materials, LTA was performed on
two regions on the much less damaged right side of the implant. These regions
were 3-4 mm apart as shown in Figure 3. One region near the edge of the device
appeared pristine (3b) while the second region was in a circa 3 x 1 mm pit that
otherwise showed no obvious damage such as discoloration (3c).
Thermal analysis in the undamaged region showed smooth and consistent LTA
curves characteristic of native UHMWPE (similar to Figure 1) but with a
Tm transition at 108°C. This lower
Tm transition, compared to the higher
Tm in Figure 1, is most likely due to lower MW
polymers that were used at the time this implant was made. In contrast to the
undamaged region, the LTA curves taken in the middle of the pit varied
substantially, with one LTA curve showing multiple transitions indicative of MW
breakdown, and a second appearing more like native PE. This variation shows that
damaged regions exhibit more varied thermal properties, likely indicative of MW,
crystallinity, oxidation, or other changes in the polymer. As noted, except for
the pit, this region otherwise appears undamaged and un-discolored by visual
inspection. Thus, thermal-mechanical analysis with the Vesta is able to detect
non-obvious changes in polymer properties.
.jpg)
Figure 3. Photograph of an explanted UHMWPE knee bearing
(a) and two regions of LTA analysis as indicated with the arrows. Region shown
in b and indicated with upper blue arrow shows no visible damage. Region c
indicated with lower red arrow is in the small pit. Local thermal analysis
curves (e) are shown for the two regions: Region b (blue curves with 108°C peak)
and region c (red curves with 118°C peaks, one of which has a second 176°C
peak.
Summary
LTA is able to discern changes in thermal properties indicative of MW,
crystallinity and/or cross-link density in conventional, radiation cross-linked,
and alpha-tocopherol impregnated UHMWPE. Additionally, LTA can detect local
changes in PE properties due to wear or other damage that are not obvious on
visual inspection, as shown with a previously implanted whole knee bearing.
Example 2: Drug Coated Cardiovascular Stents
Drug coated stents are widely used in interventional cardiology with several
hundred thousand placed in patients in the United States every year. The
distribution of the drug in the matrix polymer, the crystallinity of the drug,
and the MW and crystallinity of the polymer matrix have substantial effects on
the release kinetics, which in turn can greatly alter biological responses.
While spectroscopic microscopies such as IR and Raman can image drug
distribution within coatings at the micron scale, such methods are limited in
their capabilities in providing submicron resolution. Secondly, until now, no
known method could directly measure crystallinity or phase mixing of drugs and
matrix polymers coated onto stents. With the submicron LTA enabled with the
Vesta, this is now possible. In the following study, several drug-coated stents
were analyzed. These coatings were all based on poly-DL-lactic acid (PDLLA)
polymer matrixes, but otherwise varied in formulation.
For reasons of confidentiality the sources of the stents and their
compositions are not provided in the following table:
|
Stent Type |
Formulation |
Solvent |
|
Brand X |
PDLLA + Drug A + excipients |
Solvent 1 |
|
Brand Y |
PDLLA + Drug B + excipients |
Solvent 2 |
|
Brand Z |
PDLLA + Drug C + excipients |
Solvent 3 |
An LTA curve is shown for each stent in Figure 4, along with a
photomicrograph of the region of analysis of each stent obtained with the Vesta.
The plots show that the Tg transitions were slightly different for the three
stents, ranging from 63 to 71°C. These transitions are somewhat higher than the
typical 50-60°C Tg of pure PDLLA. This may be due to the higher
heating rate used in the LTA or due to the inclusion of the drug.
.jpg)
Figure 4. A typical LTA curve is shown for each stent
with the average peak thermal transition labeled (arrows). Micrographs of the
analyzed region are shown for each stent "brand."
The
Vesta was then used to provide Transition Temperature Maps (TTM) of Tg
measurements in an array. For each map, 36 separate LTA measurements were
automatically performed at 10 umspatial intervals in a 50 x 50 um region (Figure
5). Note the substantial differences in the uniformity and pattern of the Tg
onset for the three stents. Since the drugs have substantially higher transition
temperatures than the PDLLA matrix, this indicates that there were substantial
differences in the local content of the drug in each of the ~30 nm regions that
were thermally analyzed. For example, the Brand Y stent showed a single region
of very high drug content (as measured with Tg) surrounded by
relatively uniform composition (Tg). In contrast, there was a very
small range of drug content in Brand Z, and an intermediate range in Brand
X.
.jpg)
Figure 5. Transition temperature maps are shown for the 3
stents. For each map 36 separate measurements were obtained at 10 um spatial
intervals in an x-y grid. Each colored circle represents a single measurement,
with the onset of the Tg (°C) depicted by the color scale.
Summary
It was clearly observed that the three different stents varied considerably
in their average thermal properties, and in the morphological character of the
distribution of these thermal properties. The dominant thermal transition
observed was the Tg of the PDLLA. Overall, it has been shown that
thermal analysis performed with the Vesta can
detect significant compositional and morphological structure within drug coated
stents.
Example 3: Vestamid and Pebax Catheters
Two commercially available catheters were obtained and then examined as shown
in Figure 6. These were a catheter made from Vestamid catheter (a nylon) and
from Pebax (a block copolyamide). Compositional details of these specific
catheter samples are otherwise unknown. The Vestamid catheter showed a very
consistent Tm of 120±3°C, while the Pebax catheter showed a very
consistent Tm of 160±3°C. These Tm transitions appear
consistent with the known properties of these commercially available
materials.
.jpg)
Figure 6. LTA curves and range of melting temperature for
the Vestamid catheter (red) and the Pebax catheter (blue curves).
Summary
Vesta
analysis of these catheters showed highly uniform Tm measurements thus
indicating uniform properties.
Example 4: Contact Lenses
Two unused off-the-shelf contact lenses were examined with LTA. Both lenses
were the same size, brand and composition, but varied in power (-3.50 and
-3.00), and were produced about one year apart. These lenses were composed of a
hydrogel, but for reasons of confidentiality the source and composition are not
disclosed here. The lenses were removed from their original consumer packaging,
rinsed for over 4 hours in three 20 ml changes of distilled water to remove any
packaging salts and/or preservatives, and then air-dried overnight.
Figure 7 shows that the Tm varied slightly between the two lenses. The 3.0
and the 3.5 power lenses had respectively, mean ± standard deviation
Tm of 140.2±2°C and 144.4±1.8°C from 10 separate LTA measures of each
lens in various locations. Each lens was quite consistent in its own thermal
properties, yet the two lenses did not have the same Tm. The circa 4°C
difference in Tm between the lenses suggests that there was a small
difference in the MW of the polymers, other chemical components, and/or slight
differences in the fabrication protocols for these two similar lenses.
.jpg)
Figure 7. LTA measurement curves and the average melting
temperature (mean ± sd) of two unused off-the shelf hydrogel contact lenses. The
-3.00 lens curves are shown in red (lower set) and -3.5 lens curves are shown in
blue (upper set).
Summary
Vesta
analysis revealed uniformity of the Tm of each contact lens in 10 different
measurement locations. However, Vesta also revealed that the two different
lenses manufactured in different batches had slightly different melting
temperatures indicating differences in polymer structure or chemistry.
Conclusions
The
Anasys Vesta enables powerful analyses of thermal transition properties of
experimental and fully functional medical derive polymers. Vesta
enabled local thermal analyses (LTA) and transition temperature mapping (TTM)
reveal underling structural and compositional details that are not apparent from
visual inspection or from conventional or spectroscopic microscopy. The Vesta
analyses provide insight into drug coating structure and function, polymeric
molecular degradation, and product uniformity. Since Vesta
enables such measurements on functional fully manufactured devices, and on
explanted devices, such information can be readily applied to R&D, Quality
Control, Quality Assurance, and Failure Analysis.
Source: Nanoscale Thermal Analysis of Medical Devices
Using the VESTA
Author:Steven Goodman, Khoren Sahagian, and Kevin
Kjoller
For more information on this source please visit Anasys
Instruments