Introduction to Targeted Patch Clamping

Targeted patch clamping (TPC)1 combines patch clamping with ion conductance microscopy (ICM)2 to guide the pipette to a specific patch clamping position. Patch clamping is vital to the study of excitable cells such as neurons, cardiomyocites, and muscle fiber in electrophysiology because the technique allows the researcher to examine single or multiple ion channels in those cells. But patch clamping cannot be applied to small cells or submicron-size structures on the cell surface, due to the resolution limit of the incorporated optical microscope, which is used to maneuver the patch clamp’s pipette near the cell surface. Because ICM can identify the cell surface at the submicron scale using the same patch clamping pipette, TPC overcomes the optical resolution hurdle, and greatly improves the applicability of the patch clamp and its accuracy beyond the classical technique. In this paper, live rat ventricular cardiomyocyte cells were examined with targeted patch clamping using a Park ICM. Ion channel signals were successfully recorded at a chosen Z groove location on the ventricular cardiomyocyte.

Operational Principle of Targeted Patch Clamping

The main idea of TPC is to perform patch clamping on small cellular structures using ICM for submicron-scale detection. The ICM utilizes the same pipette and electric circuit used with traditional patch clamping to detect the cell topography using Park XEP control software in the ICM imaging mode, high resolution cell topography is acquired with feedback control. Once an interesting structure for patch clamping is identified via ICM imaging, the ICM Z scanner positions the pipette on the structure. A giga-ohm seal is formed by applying suction. Ion channel recording can then be performed as in conventional patch clamping. Because the pipette approaches the cell surface in the vertical direction with nanometer scale precision, TPC improves the probability of making agiga-ohm seal compared with the slanted approach used in conventional patch clamping3.

Figure 1. Illustration of the operational steps for targeted patch clamping: (a) Park ICM acquires the topography of the cardiomycyte cell surface. (b) The nano-pipette is positioned on the chosen region (z groove) by the XY scanner of the ICM. (c) The nano-pipette approaches and forms a giga-ohm seal to record the channel opening current signal.

Instrumentation for Targeted Patch Clamping

To implement TPC, the ICM is integrated with a patch clamp system by sharing the pipette and its ionic current circuit. ICM receives the ionic current signal for imaging from the patch clamp's ion current detector (pre-amp.). The ionic current flows through the nano-pipette, it is detected by the patch clamp's amplifier, and then it is divided into two: one for the ICM cell imaging (feedback control) and the other for the electrophysiological recording.

Figure 2. Configuration of targeted patch clamping: (a) XE-Bio ICM integrated with the Axon Axopatch 200B. (b) A schematic diagram of patch clamp connection with XE-Bio ICM.

Samples and Methods

Live rat ventricular cardiomyocyte cells were isolated as previously described4. Cells were allowed to attach to polystyrene cell culture dishes filled with bath solutions composed of 120 NaCl, 5.4 KCl, 5 MgSO4, 0.2 CaCl2, 5 Na-Pyryvate, 5.5 Glucose, 20 Taurine, 10 Hepes, and 29 Manitol, adjusted to PH 7.4 with NaOH. Pipette filling solution was the same as the bath solution. A live cell image was acquired in ICM-ARS mode5. The pipette used for the ICM was pulled from borosilicate glass (O.D.,1.0 mm, I.D., 0.58 mm, length 90 mm; Warner Instruments, US) using a CO2-laser-based micropipette puller (P-2000, Sutter Instruments, Novato, US). The inner diameter of the pipette was about 500 nm.

Figure 3. (a) Optical image (top inset) and ICM topography of a live rat ventricular cardiomyocyte. The small red circle indicates a Z groove structure where patch clamping was performed. (b) Current transient (3.45 pA) of cell attached, recording at voltage of ±0 mV. (c) The conductance measurement of the recorded channel. The conductance is 44 pS, and was estimated to be the inward-rectifier K channel.

Ion Channel Recording of Targeted Patch Clamping

Figure 3 shows an example of targeted patch clamping using ICM. ICM topography illustrates typical Z grooves clearly, with details. At the position on the Z groove structure (red circle) selected from the ICM topography image, we examined the attached cell recording and successfully measured a transient current of about 3.45 pA at a voltage of ±0 mV. This result indicates that Park XE-Bio ICM is well-suited for both precise targeting and patch clamping on the cell membrane. The ICM is especially useful for research on small cells such as sperm cells, subcellular structures such as microvilli6, and even opaque cells like intact tissues, trachea, and brain slices.

References

1. J. Gorelik, G. Yuchun et al. Biophys. J. 83, 3296 (2002)
2. P. K. Hansma, B. Drake, O. Marti, S. A. Gould, C. B. Prater, Science 243, 641 (1989)
3. O.P. Hamill, A. Marthy, E. Neher, B. Sakmann, F.J. Sigworth, J. Physiol. 391, 85 (1981)
4. S.E. Harding, G. Vescovo, M. Kirby, S.M. Jones, J. Gurden, and P.A. Poole-Wilson, J.Mol. Cell. Cardiol. 20. 635 (1988).
5. C. Myunghoon, XE-Bio Application Note 1, Park Systems (2012)
6. G. Yuchun, G. Julia, et al. FASEB J. 16, 748 (2002)

Authors

Myung Hoon Choi, Goo Eun Jung (Research Product Management, Research & Development of Park Systems, Seoul, Korea)

About Park Systems

Park Systems is the leading nanotechnology solutions partner for the most challenging problems of nanoscale research and industrial applications.

Park Systems provides original and innovative AFM solutions for the most accurate nanoscale measurement. In nanoscale metrology, having data that is repeatable, reproducible, and reliable is just as crucial as achieving high-resolution. The innovative crosstalk-elimination (XE) metrology platform ushered in a new era of nanometrology that overcomes non-linearity and non-orthogonality associated with conventional piezotube based systems. Park Systems innovative AFM technology is a disruptive market force and it expands the application of nanometrology beyond the limits of conventional AFM technology.

This information has been sourced, reviewed and adapted from materials provided by Park Systems.

For more information on this source, please visit Park Systems.