By AZoNano
Table of Contents
Introduction
Features of AcuityXR
Case 1: Improved Metrology of Narrow Linewidths
Case 2: Feature Separation Capability
Case 3: General Feature Clarity
Limitations of AcuityXR Mode
Conclusions
About Bruker
Introduction
Bruker's
AcuityXR optical surface profiling system has been designed in such a way that it can significantly
enhance the lateral resolution for a broad class of measurements. AcuityXR
works on any smooth surface in which the phase of the light is examined
and is used to calculate the surface from the
white-light-interferometric signal.
Features of AcuityXR
The key features of the AcuityXR
are:
- AcuityXR produces a measurement with twice the
number of pixels in the X and Y directions than is possible with a
standard interferometric measurement.
- Greater detail is achieved for fine features in the measurement,
such as more separated lines and sharper scratches or defects. There
may be slight ringing near the edges of the field of view or near areas
of missing surface data from the processing needed to improve
resolution, but this is a normal consequence of the mathematical
algorithms used to enhance the resolution.
The examples below detail several use cases where AcuityXR
measurements provide significant benefits in detection capability or in
quantification of key surface features.
Case 1: Improved Metrology of Narrow Linewidths
One obvious application of improved lateral resolution is to attain
superior measurement of narrow linewidth features. To test this, a
sample having two 350nm lines, separated by 350nm were measured on a
Bruker ContourGT X8 Optical Profiler System using standard Phase
Shifting Interferometry (PSI) and AcuityXR
PSI measurement 3 modes. Figure 1 shows one measurement from each
technique.
The observations are listed below:
- First, it can be clearly observed that by utilizing the AcuityXR
measurement, sufficient resolution is reached to show the full
separation between the lines, whereas for the PSI image they are
blurred enough such that the surface height between the lines is not
the same as the rest of the substrate.
- Additionally, fine details of the linewidth variation along the
line are seen in the AcuityXR
result and not apparent in the standard PSI result. The average
linewidth determined using PSI was just 300 nm. Using AcuityXR
mode, the average linewidth was 320 nm. Though the AcuityXR
result is still slightly smaller than listed on the standard, it can be
adjusted based on exactly where the width is taken vertically along the
line.
- In terms of numerical measurement performance, the standard
deviation of line width for the PSI measurements was 35.1 nm. For the AcuityXR
PSI results, the standard deviation is reduced by nearly a factor of 7,
to just 5.2 nm. Similarly, during calculation of the line volume, the
standard deviations for PSI and AcuityXR
PSI measurements are .005 mm3 and 0.001 mm3 respectively, which is a factor of 5
improvement. Hence, significant gains in metrology capability are
possible through the enhanced lateral resolution offered by AcuityXR.

Figure 1. 350nm linewidth measurements taken
with Standard PSI (left) with little feature separation and AcuityXR
PSI (right), showing high levels of feature differentiation.
Case 2: Feature Separation Capability
In order to actually test the limits of AcuityXR
and it’s ability to differentiate fine features, a linewidth standard
of successively smaller lines, from 1mm to
100nm, was obtained. Bruker’s exclusive 115X, 0.8NA microscope
objective was utilized to offer maximum resolution on the linewidth
standard. The lateral resolution of the optical system as obtained
using the Sparrow criterion is 314 nm. Features up to 200nm can be
easily differentiated both using PSI and AcuityXR
PSI modes as shown in Figure 2. However, finer detail and a superior
separation are clearly observed using the AcuityXR
mode.
At 130nm there was no separation at all visible using normal PSI
measurement, while AcuityXR
mode was able to show two distinct features. This apparent limit of the
lateral resolution capability of AcuityXR
enhanced measurements is around 2.5 times smaller than the lateral
resolution anticipated from the Sparrow calculation.

Figure 2. All the images have the same
vertical scale of 2nm (red) to -8nm (blue). Standard PSI Measurements
(left) and AcuityXR PSI Measurements (right) of 200, 150, and 130nm
linewidth features.
Case 3: General Feature Clarity
Normally when examining surfaces, it is essential to identify
scratches, defects, or other fine features without necessarily needing
to perform gage studies or otherwise quantify each individual feature.
For example, this would be a nonsensical laboratory requirement when
examining surface finish in orthopedics, medical implants, plastic
substrates, or optics. A variety of surfaces were rapidly measured
using PSI measurement mode and AcuityXR
PSI mode as shown in Figure 3.

Figure 3. Images taken with Standard PSI
(left) and AcuityXR PSI (right) shows vast improvement in AcuityXR PSI
in the ability to make images less pixilated while showing the proper
structure on the sample.
In each case, the AcuityXR
measurement shows sharper features on the gratings and other small
defects present in the images. This allows not just superior visual
identification, but also a better capability for automatically
locating, counting, and quantifying features based on width or height
with Bruker’s Vision software.
Limitations of AcuityXR Mode
The limitations of AcuityXR
are listed below:
- It is limited to surfaces having local roughness less than
approximately 20nm, where PSI or HDVSI measurement modes can conduct
measurement of the surface without problems.
- Rough surfaces, such as paper, roughly machined metal, or foams
are not appropriate for AcuityXR.
- Also, surfaces having large amounts of missing data may suffer
some ringing near the edges of the valid data when using AcuityXR,
since the mathematics involved require contiguous surfaces for proper
calculation.
Conclusions
AcuityXR
is an innovative technology available for most models of Bruker’s
optical profilers. It employs system modeling, low-noise measurements,
and the integration of multiple surface scans. With this integration,
the blur caused by the optical elements can be reduced and lateral
resolution considerably enhanced. Greater detail can be seen in many
surfaces. For narrow features, AcuityXR
also offers significantly enhanced quantification of variations, making
process control possible even on small structures. While AcuityXR
is not suitable for all surfaces, for smooth, fine features it improves
the measurement capability of the optical profiler.
About Bruker
Bruker
Nano provides Atomic Force Microscope/Scanning Probe Microscope
(AFM/SPM) products that stand out from other commercially available
systems for their robust design and ease-of-use, whilst maintaining the
highest resolution. The NANOS measuring head, which is part of all our
instruments, employs a unique fiber-optic interferometer for measuring
the cantilever deflection, which makes the setup so compact that it is
no larger than a standard research microscope objective.

This information has been sourced, reviewed and
adapted from materials provided by Bruker AXS.
For more information on this source, please visit Bruker
AXS.