|
Polymer latex spheres are very commonly used
to verify the operation of particle size analysers. This is because they are
available as monosize dispersions of near perfect spheres. The sphere is the
only three-dimensional shape whose size can be unambiguously described by a
single figure and being monodisperse removes any uncertainty regarding the
calculation of a mean size. Polymer latices have other benefits. They have a
similar density to water, so particles less than 1 micron will remain in
suspension during measurement. Dispersions can be stored at room temperature and
have storage lifetimes of months or years.
Monodisperse
Lattices for Calibration
A wide range of monodisperse polystyrene
lattices are available from a variety of manufacturers, however not all are
supplied with an individual calibration certificate. Duke scientific corporation
is one manufacturer that supplies a calibration certificate with each sample,
measured by transmission electron microscope, (TEM) which is traceable to NIST
certified microspheres. The specification for the standards also includes a
hydrodynamic diameter measured by photon correlation spectroscopy
(PCS).
Standards for Photon
Correlation Spectroscopy
Standards suitable for PCS are available
from 20nm to 900 nm. The easiest sizes to measure are in the range 60nm to
300nm. Particles larger than 60nm are large enough to give very reproducible
results with low power lasers at suitable dilutions for PCS. Particles larger
than 300nm start to show a marked variation in scattering intensity with angle,
measuring standards smaller than this removes the requirement to consider the
angle.
Certified Size
The result quoted on the Duke latex bottle
is the certified TEM result. The PCS result is quoted in the specification sheet
and is not a certified value. Table 1 lists a few comparisons of these two
figures. For all Duke standards the size accuracy by PCS should be within the
specified PCS range ±2% for samples prepared in a 1mM salt such as NaCl or
KNO3. This figure should account for uncertainties in sample
preparation. Measurement precision should be ±1% or better. The peak width is
expressed as the polydispersity. For polydispersities less than 0.2, this is
equivalent to the variance of the distribution. A standard latex correctly
dispersed should have a polydispersity less than 0.03.
Table 1. Quoted sizes from Duke
Scientific Corp. Polystyrene latex standards specification sheets.
|
|
|
|
|
|
15586 |
JUL/1/94 |
105±3 |
105-112 |
|
15329 |
APR/18/94 |
149±4 |
150-156 |
|
15504 |
JUN/7/94 |
220±6 |
220-227 |
|
14919 |
DEC/7/93 |
269±7 |
267-275 |
Comparison of Sizes
Measured by TEM and PCS
It is easy to forget that different
measurement techniques measure different properties of a particle and so can
give different results. The question often arises, which is the correct
result?
Electron Microscopy
versus Photon Correlation Spectroscopy
For many people ‘seeing is believing’ so the
electron microscope result is ‘correct’. In fact, samples prepared for electron
microscope examination are often harshly treated, this treatment can distort
soft materials such as polymer lattices and change or mask surface structures.
It can make the size measurement of some types of materials like surfactant
micelles impossible. PCS in contrast measures the Hydrodynamic diameter of
dispersed particles in their native environment.
Factors that can Affect the
Measured Size or a Polymer
Any surface structure such as a ‘hairy’
surface made up of polymer chains, or a change in the electrical double layer
that affects the Brownian motion of the particle, will change the effective
particle size. Increasing the surface structure or extending the double electric
layer by using a very low salt dispersant, will reduce Brownian motion and
increase the measured size.
For these reasons, the hydrodynamic size or
PCS size of particles that are not smooth hard spheres, is usually larger than
the TEM size.
Sample Preparation For
Measurement By PCS
All latex standards are supplied at a
concentration that is too high for PCS measurement, typically 2% w/v. The latex
dispersion should be diluted with distilled, or preferably demineralised water
filtered down to 0.2 microns.
Measurements in Salt
Solutions
The final concentration should be such that
the result is independent of the actual concentration, 0.002%w/v is a good
guide, but the optimum concentration will be size dependent. A practical
criterion, if a spectrophotometer is available, is that the optical density in a
1cm cell should be less than 0.04. If a size is required that more nearly
matches the microscope result, a dilute simple salt such as NaCl or KCl should
be used instead of pure demineralised water. This will compress the double
electric layer and reduce the effective size. Table 2 shows a comparison of
sizes in different salt concentrations.
Table 2. Comparison of measured sizes in
various salt solutions.
|
|
|
|
|
TEM result (nm) |
105±3 |
220±6 |
|
PCS Result (nm) |
|
in demin water |
114.6pd. 0.01 |
235.1 pd 0.01 |
|
in 1mM NaCl |
107.7 pd 0.01 |
222.3 pd 0.01 |
|
in 10mM NaCl |
107.5 pd 0.02 |
226.1 pd 0.02 |
|
Samples measured 72 hrs after
dilution |
|
in demin water |
114.1 pd 0.01 |
233.4 pd 0.02 |
|
in 1mM NaCl |
107.0 pd 0.01 |
221.8 pd 0.02 |
|
in 10mM NaCl |
108.5 pd 0.05 |
231.1 pd 0.05 |
The 105nm latex was prepared at 0.001% w/v,
the 220nm latex at 0.0005% w/v.
All measurements were done at an angle of
90° and a temperature of 28°C with a 488nm laser set at 30mW. The analysis time
was 300 s. The analysis was set to monomodal.
Troubleshooting
There are some simple settings to check if
the result obtained is not as expected.
Temperature
Equilibration and Calibration
If the sample temperature is input or read
incorrectly the viscosity will be calculated incorrectly and the reported size
will be wrong. For aqueous systems such a latex dispersion, at 20°C a 1 degree
error in the temperature will result in a 2.4% error in the viscosity used and
therefore a 2.4% error in the calculated size.
Temperature equilibration can take several
minutes if the sample temperature has to change by more than 2 or 3 degrees.
Doing several measurements is a good check that the temperature is
stable.
Viscosity
This is calculated from the temperature for
aqueous systems. To ensure this calculation is being done, enter 0 (zero) for
the viscosity on the measure document page.
Dispersant
Refractive Index
The refractive index of the continuous
phase, 1.330 for pure water.
Wavelength
The wavelength of the laser used in
nanometers. 633 for a Helium Neon laser, generally 488 or 515 for an Argon ion
laser.
Angle
For a variable angle system, check that the
angle displayed on the measure document page is the same as the actual angle of
the spectrometer.
Measurement Time
Reasonable results can be obtained in a few
seconds, but a measurement time of over 100 seconds will ensure automatic dust
rejection during data collection and therefore a more reproducible result.
Setting the measurement time to auto will ensure sufficient quality data is
collected for a reliable measurement of standard latices.
Summary
|
|
|
|
|
Too large or small |
Less than 0.03 |
Temp. calibration measurement setting |
|
Too large |
Greater than 0.1 |
Contaminated dilutent, measurement time too
short |
|
Too small |
Less than 0.03 |
Salt concentration too low |
|
Drifting |
Greater than 0.03 |
Concentration too high |
|
Contrast |
Greater than 0.03 |
Equilibration time too short |
|
Correct |
Greater than 0.03 |
Measurement time too short |
Conclusions
- Measure latex standards in a 10mM salt
solution, such as sodium chloride in demineralised water, rather than in pure
water.
- Dilute the samples on the day of
measurement.
- Determine the correct sample concentration
for each sample and set of measurement conditions.
- Set all measurement parameters to ‘Auto’.
Note: A complete set of references can be found
by referring to the source document.
Source: "The Measurement of Latex Standards by Photon
Correlation Spectroscopy”, Application Note by Malvern
Instruments.
For more information on this source please
visit Malvern
Instruments Ltd (UK) or Malvern Instruments
(USA).
Date Added: May 12, 2005
|