Thermo Scientific™ Evolution™ UV-Vis Spectrophotometers integrate dependable and flexible hardware with intuitive, user-friendly Thermo Scientific™ Insight™ Pro Software.
Together, they meet the diverse needs that any organization requires from an analytical tool.
Product Overview
- With intuitive INSIGHT Pro Software supporting each stage of the workflow, users can quickly progress from sample to final report.
- Workflow-focused application modules simplify the customization of even the most complex procedures.
- With comprehensive software tools for data collection, processing, and reporting, users can obtain the required results for quantitative analysis, scanning, and kinetics applications.
- Optional INSIGHT Pro Security Software features help meet current 21 CFR Part 11 compliance requirements in the pharmaceutical laboratory by ensuring data security.
Enhance Productivity with Sampling Flexibility
- Thermo Scientific™ Smart Accessories feature a snap-in design that ensures ease of use and consistency, enabling customers to complete analysis more quickly.
- Auto-recognition and software integration of sippers, cell changers, and autosampler accessories remove the need for manual setup.
- The Evolution Pro UV-Vis Spectrophotometer’s large sample compartment and wide beam separation support a range of accessories and innovative applications.
Regulatory/Quality Control Environments
A range of compliance solutions and services is offered to support users in operating the Evolution Pro UV-Vis Spectrophotometer in regulated or quality-control environments, minimizing implementation time and maintaining system performance.
Evolution Pro Pharma Value Bundles Include
- Qualification Filter Kit for UV-Vis
- INSIGHT Pro with Security Software
- Validator for Insight Pro US Ltr or INTL A4 Paper Version
- Evolution Pro Spectrometer
Evolution Pro USP/EP Standards Pharma Bundles Include
- USP and EP Visible Standards Set
- Evolution Pro Spectrometer
- Validator for Insight Pro US Ltr or INTL A4 Paper Version
- USP and EP UV Standards Set
- INSIGHT Pro with Security Software
Accessory Ordering Information
All accessories can be purchased individually. Additional options include cell changers, calibration carousels, cuvette and test holders, sippers, small-volume sampling tools, fiber optic probes, and accessories for rapid mixing and reflectance.
Download the Brochure for More Information
Specifications
Source: Thermo Fisher Scientific - UV-Vis Spectroscopy
| Specifications |
| Baseline Flatness |
± 0.001 A (200-800 nm), 2 nm SBW, smoothed |
| Drift |
< 0.0005 A&fraslhr., 500 nm, 2 nm SBW, 2-hour warmup |
| Electrical Requirements |
100/240 V, 50/60 Hz |
| For Use With (Application) |
Academic, Chemicals, Food and Beverage, Chemicals Environmental, Industrial QA/QC, Material Science, Pharmaceuticals |
| Green Features |
ACT Ecolabel (Gold) |
| Includes |
Evolution Pro UV-Vis Spectrophotometer |
| Lamp Life |
Five years or longer if not using a live signal |
| Noise |
0 A: < 0.00018 A 1 A: < 0.00022 A 2 A: < 0.00050 A 500 nm, 2 nm SBW, RMS |
| Optical Design |
Modified Ebert Double beam with sample and reference cuvette/accessory positions |
Pharmacopoeia Compliance Testing |
Wavelength Accuracy: ± 0.3 nm 190–900 nm Wavelength Repeatability: SD 10 meas. < 0.1 nm Absorbance Accuracy: < ± 0.010 A at 1 A, < ± 0.010 A at 2 A Absorbance Repeatability: ± 0.0025 A @ 1 A Stray Light: |
| Photometric Accuracy |
1 A: ± 0.004 A, 2.0 A: ± 0.004 A, 3.0 A: ± 0.006 A, Measured at 440 nm using neutral density filters traceable to NIST |
| Photometric Display |
± 6 A |
| Photometric Range |
> 4 A |
| Photometric Repeatability |
1 A: ± 0.0025 A |
| Scan Ordinate Modes |
Absorbance, % Transmittance, % Reflectance, Kubelka-Munk, Log(1/R), Log(Abs), ABS × Factor, Intensity, first to fourth Derivative |
| Scan Speed |
Variable, up to 6000 nm/min. |
| Stray Light |
KCl, 198 nm: ≤ 0.4 % T NaI, 220 nm: ≤ 0.032 % T NaNO2, 340 nm: < 0.01 % T |
| Warranty |
Three-year Source Replacement Warranty |
| Wavelength Accuracy |
± 0.20 nm (546.07 nm Hg emission line), ± 0.30 nm, 190 to 900 nm |
| Wavelength Data Interval |
10, 5, 2, 1, 0.5, 0.2, 0.1, 0.05 nm |
| Wavelength Repeatability |
Standard deviation of 10 measurements < 0.05 nm |
| Detector Type |
Dual-Matched Silicon Photodiodes |
| Light Source |
Xenon Flash |
| Product Line |
Evolution Pro |
| Spectral Bandwidth |
0.5, 1.0, 1.5, 2.0, 4.0 nm selectable |
| Type |
UV-Vis Spectrophotometer |
| Wavelength Range |
190-1100 nm |
| Unit Size |
Each |