Posted in | Spectrophotometers

Evolution Pro UV-Vis Spectrophotometers for Laboratory Analysis

Thermo Scientific Evolution UV-Vis Spectrophotometers integrate dependable and flexible hardware with intuitive, user-friendly Thermo Scientific Insight Pro Software. 

Together, they meet the diverse needs that any organization requires from an analytical tool. 

Product Overview

  • With intuitive INSIGHT Pro Software supporting each stage of the workflow, users can quickly progress from sample to final report.
  • Workflow-focused application modules simplify the customization of even the most complex procedures.
  • With comprehensive software tools for data collection, processing, and reporting, users can obtain the required results for quantitative analysis, scanning, and kinetics applications.
  • Optional INSIGHT Pro Security Software features help meet current 21 CFR Part 11 compliance requirements in the pharmaceutical laboratory by ensuring data security.

Enhance Productivity with Sampling Flexibility

  • Thermo Scientific Smart Accessories feature a snap-in design that ensures ease of use and consistency, enabling customers to complete analysis more quickly.
  • Auto-recognition and software integration of sippers, cell changers, and autosampler accessories remove the need for manual setup.
  • The Evolution Pro UV-Vis Spectrophotometer’s large sample compartment and wide beam separation support a range of accessories and innovative applications.

Regulatory/Quality Control Environments

A range of compliance solutions and services is offered to support users in operating the Evolution Pro UV-Vis Spectrophotometer in regulated or quality-control environments, minimizing implementation time and maintaining system performance.

Evolution Pro Pharma Value Bundles Include

  • Qualification Filter Kit for UV-Vis
  • INSIGHT Pro with Security Software
  • Validator for Insight Pro US Ltr or INTL A4 Paper Version
  • Evolution Pro Spectrometer

Evolution Pro USP/EP Standards Pharma Bundles Include

  • USP and EP Visible Standards Set
  • Evolution Pro Spectrometer
  • Validator for Insight Pro US Ltr or INTL A4 Paper Version
  • USP and EP UV Standards Set
  • INSIGHT Pro with Security Software

Accessory Ordering Information

All accessories can be purchased individually. Additional options include cell changers, calibration carousels, cuvette and test holders, sippers, small-volume sampling tools, fiber optic probes, and accessories for rapid mixing and reflectance.

Download the Brochure for More Information

Specifications

Source: Thermo Fisher Scientific - UV-Vis Spectroscopy

Specifications
Baseline Flatness ± 0.001 A (200-800 nm), 2 nm SBW, smoothed
Drift < 0.0005 A&fraslhr., 500 nm, 2 nm SBW, 2-hour warmup
Electrical Requirements 100/240 V, 50/60 Hz
For Use With (Application) Academic, Chemicals, Food and Beverage, Chemicals Environmental, Industrial QA/QC, Material Science, Pharmaceuticals
Green Features ACT Ecolabel (Gold)
Includes Evolution Pro UV-Vis Spectrophotometer
Lamp Life Five years or longer if not using a live signal
Noise 0 A: < 0.00018 A
1 A: < 0.00022 A
2 A: < 0.00050 A
500 nm, 2 nm SBW, RMS
Optical Design Modified Ebert
Double beam with sample and reference cuvette/accessory positions
Pharmacopoeia
Compliance Testing
Wavelength Accuracy: ± 0.3 nm 190–900 nm
Wavelength Repeatability: SD 10 meas. < 0.1 nm
Absorbance Accuracy: < ± 0.010 A at 1 A, < ± 0.010 A at 2 A
Absorbance Repeatability: ± 0.0025 A @ 1 A
Stray Light:
Photometric Accuracy 1 A: ± 0.004 A, 2.0 A: ± 0.004 A, 3.0 A: ± 0.006 A, Measured at 440 nm using neutral density filters traceable to NIST
Photometric Display ± 6 A
Photometric Range > 4 A
Photometric Repeatability 1 A: ± 0.0025 A
Scan Ordinate Modes Absorbance, % Transmittance, % Reflectance, Kubelka-Munk, Log(1/R), Log(Abs), ABS × Factor, Intensity, first to fourth Derivative
Scan Speed Variable, up to 6000 nm/min.
Stray Light KCl, 198 nm: ≤ 0.4 % T
NaI, 220 nm: ≤ 0.032 % T
NaNO2, 340 nm: < 0.01 % T
Warranty Three-year Source Replacement Warranty
Wavelength Accuracy ± 0.20 nm (546.07 nm Hg emission line), ± 0.30 nm, 190 to 900 nm
Wavelength Data Interval 10, 5, 2, 1, 0.5, 0.2, 0.1, 0.05 nm
Wavelength Repeatability Standard deviation of 10 measurements < 0.05 nm
Detector Type Dual-Matched Silicon
Photodiodes
Light Source Xenon Flash
Product Line Evolution Pro
Spectral Bandwidth 0.5, 1.0, 1.5, 2.0, 4.0 nm selectable
Type UV-Vis Spectrophotometer
Wavelength Range 190-1100 nm
Unit Size Each

Other Equipment by this Supplier

Other Equipment

While we only use edited and approved content for Azthena answers, it may on occasions provide incorrect responses. Please confirm any data provided with the related suppliers or authors. We do not provide medical advice, if you search for medical information you must always consult a medical professional before acting on any information provided.

Your questions, but not your email details will be shared with OpenAI and retained for 30 days in accordance with their privacy principles.

Please do not ask questions that use sensitive or confidential information.

Read the full Terms & Conditions.